1. From the Summary screen, tap the Calibration button.

2. From this screen you can tap the Learn More button to access an FAQ about calibration. The screen will also provide information of the last time and how the sensor was calibrated. To begin the calibration process, tap the Calibrate button.

3. From this screen, you can choose which calibration technique you want to use by tapping either the 1 Point or 2 Point buttons. If you do not have a Boveda Calibration Kit, you can purchase one by tapping the Buy One button. For a 2 Point calibration, tap 2 Point.

4. Instructions are provided on how to setup the calibration, but for more information on using the kit, tap Learn More. In a 2 Point Calibration, the first calibration point will be performed at 75% RH. Once the sensor is in the sealed 75% RH calibration kit bag, tap I'm Ready, Start.

5. The calibration has now started and must continue for a full 24 hours.

6. As the calibration progresses, the App counts down the time remaining. The App will let you know when the calibration has completed.

7. When the calibration at 75% RH completes, instructions will be displayed to place the sensor in the 32% RH calibration kit bag. Once the sensor is in the sealed 75% RH calibration kit bag, tap I'm Ready, Start.

8. The calibration has now started for the second calibration point and must continue for another full 24 hours.

9. As the calibration progresses, the App counts down the time remaining. The App will let you know when the calibration has completed.

10. When the calibration finishes, tap the OK button.

11. The calibration Summary screen will appear next showing the information for the calibration. To view the offset applied to the RH, tap the Customize button.

12. The High Point and Low Point Offsets will be displayed. Tap the Back button in the upper left hand corner of the screen to exit calibration.